When you have finished changing all the parameter units and count time settings you want to
change, you may
1. Press [F1] to select another setting group you want to change; or;
2. Press [MENU] and a message box will pop up to remind you to save the changes, as
Figure 5-32 shows. CLICK "Yes" to save the changes and exit to the system menu; or
CLICK "No" to exit to the system menu without saving the changes.

5.3.6 Reference Range

You can set a reference range for every parameter. The system will flag any analysis result
that exceeds this range with either an H or L. This analyzer divides the patients into 5 patient
groups, which are all listed in Table 5-3. The default ranges are given. Note that the default
ranges are for references only. Each laboratory is encouraged to establish its own
references.
Group
General
Not specified, male or female.
Not specified.
Man
Male
Woman
Female
Child
Male or Female
Neonate
Male or Female
The upper and lower limits of the reference ranges are visible to all users but changeable
Customizing the Analyzer Software
Figure 5-32 Saving changes
Table 5-3 Demographic groups
Gender
5-20
Age
Not specified.
> 12 years
> 12 years
> 12 years
> 28 days and ≤12 years
≤ 28 days