Mindray BC-2800 Operator's Manual page 278

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

Message description unit
Field
name
MD
Programming
The BC-2800 must communicate with the external computer when the Handshake is on in
following procedures:
1. During the communication connection, the BC-2800 sends an ENQ (10 Hex), then waits
up to 4 seconds for the external computer to respond. If the external computer does not
respond, then one more ENQ (10 Hex) is tried. If it fails again, the analyzer aborts the
transmission and reports a transmission error;
2. The external computer must respond by sending an ACK (06 Hex);
3. After the ACK (06 Hex) is received, the analyzer then sends data block;
4. After the sending, the BC-2800 sends an ETX (0F Hex), then waits 4 seconds for the
external computer to respond. If no response is received, one more ETX (0F Hex) is sent.
The BC-2800 waits 4 seconds before giving up and gives alarm of communication error.
E.3.2 Sample Data format
FD field
FD
Sample information
SampleID
Mode
TestTime
Name
Appendices
Coding
CTR
results
QCR
QCC
FD definition of analysis parameter
Description
Sample ID
Analysis Mode (0:whole blood-all parameter; 1: predilute-all parameter;
2: whole blood- WBC/HGB; 3: predilute - WBC/HGB; 4: whole
blood-RBC/PLT; 5: predilute - RBC/PLT)
Test time, format: YYYY-MM-DD HH: MM: SS (if the digit is less than
10, add a 0 at the ten's place)
Name
E-4
and V
Description
Data package of normal analysis
Data package of QC running
Data package of QC standard