10. When the analysis is finished, the result will be displayed on the screen and the sample
probe will be repositioned.
Keep dust from the prepared diluent.
Evaluate predilute stability based on your laboratory's sample population
and sample collection techniques or methods.
If the analyzer detects WBC/RBC clogging or bubbles during the analysis,
the corresponding error messages will be displayed and the results of all the
related parameters will be invalidated. See Chapter 11 Troubleshooting Your
Analyzer for solutions.
If the ambient temperature is outside the specified operating range, the
analyzer will alarm you for abnormal ambient temperature and the analysis
results may be unreliable. See Chapter 11 Troubleshooting Your Analyzer for
solutions.
Saving the calibration results
If non-numeric parameter values ("***") are obtained, a message box will pop up to warn you,
as Figure 9-6 shows. CLICK "Yes" to close the dialog box and discard the result.
Figure 9-6 A message box to warn you about the invalid results
If the parameter values obtained are numeric, a message box will pop up to confirm the
validity of the results, as Figure 9-7 shows.
Using the Calibration Programs
9-8