Maintaining Your Analyzer
Figure 10-38 Testing valves
To test a valve, SELECT the valve. If the valve goes through an Off-On-Off sequence without
making any abnormal sound, it passes the test. Otherwise, something may be wrong with the
valve.
10.5.4 Testing A/D Interrupt
To test the A/D interrupt, press [F1] to select the "Circuit" group, as Figure 10-39 shows.
Figure 10-39 Testing A/D interrupt
To conduct the test, SELECT "A/D interrupt" and the test result will be displayed later.