Mindray BC-2800 Operator's Manual page 159

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

Editing QC settings
If there are saved QC results and settings, you need to delete them first. You can press [F5]
to enter "QC Table" screen to delete all the results, see Chapter 8.2.3 for details.
Entering the "QC Edit" screen
Press [F3] at "Controls" screen to enter the "QC Edit" screen (Figure 8-3).
Entering lot number
ENTER the lot number of the control to be used into the "Lot No." box, as Figure 8-4 shows.
Entering expiration date
ENTER the expiration date of the control to be used into the "Exp. Date" box, as Figure 8-5
Using the QC Programs
Figure 8-3 "QC Edit" screen
Figure 8-4 Entering lot number
8-3