Mindray BC-2800 Operator's Manual page 77

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

For example, assuming the expected MCV result is 90.0fL, while the actual analysis result is
82.0fL, then
ExpectedMC
You should adjust the RBC gain to 109.8% as close as possible. Follow the steps given below
to do so.
1. Enter the administrator password as introduced in Chapter 5.2.1;
2. At the "Settings" screen, press [F1] to select the "Gain" group;
3. SELECT "RBC", as Figure 5-26 shows;
4. ENTER the desired gain so that the adjustment becomes as close to 109.8% as possible.
Setting the HGB gain
You may adjust the HGB gain to change the HGB blank voltage, which usually should be set
between 3.4 to 4.8V (4.5V recommended). Follow the steps given below to set the HGB gain.
1. Enter the administrator password as introduced in Chapter 5.2.1;
2. At the "Settings" screen, press [F1] to select the "Gain" group;
3. SELECT "HGB";
4. ENTER the desired gain so that the HGB blank voltage is between 3.4 to 4.8V, as Figure
5-27 shows.
Customizing the Analyzer Software
V
×
100
%
ActualMCV
Figure 5-26 Setting RBC gain
90
0 .
×
=
100
%
109
82
0 .
5-15
8 .
%