Fresh Blood Calibration - Mindray BC-2800 Operator's Manual

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

Verifying new calibration factors
Press [MENU] to exit the "Calibrator" screen, a message box will pop up to confirm the new
calibration factors, as Figure 9-9 shows.
Figure 9-9 A message box to confirm the new calibration factors
CLICK "Yes" to save the new calibration factors and enter the "Count" screen from the
system menu.
At the "Count" screen, run the calibrator or a normal-level control at least 3 consecutive times
and calculate the mean of the results. Compare the obtained means to the expected means.
Other operations
Printing new calibration factors
Press [PRINT] to print out the current calibration factors.
Exiting the "Calibrator" screen
Press [MENU] to exit to system menu.

9.3.3 Fresh Blood Calibration

ENTER the administrator password in the "Password" screen. Press [MENU] to enter the
system menu.
Using the Calibration Programs
Figure 9-10 System menu
9-10