should be no less than 8cm, below the sample probe;
7. Press [ENTER] to soak the probe wipe with the aspirated cleanser. The soaking progress
will be displayed on the screen, as Figure 10-27 shows;
8. When the soaking is done, wipe the bottom of the probe wipe with a probe
cleanser-dipped cloth that does not leave debris;
9. Press [ENTER] to flush the block and the interior of the probe and the flushing progress is
displayed on the screen, as Figure 10-28 shows;
Figure 10-28 Flush the block and the interior of the probe
10. After the flushing is done, the screen returns to the initial state.
When the accumulated analyzed samples reach 150, a message box will pop up to remind to
Maintaining Your Analyzer
Figure 10-27 Cleaning probe wipe
10-20