Mindray BC-2800 Operator's Manual page 244

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

Recorder head too hot.
Recorder too hot
Insufficient lyse or wrong lyse
Lyse out
volume setting.
Expired diluent or wrong
Diluent expired
expiration setting
Expired rinse or wrong
Rinse expired
expiration setting
Expired lyse or wrong
Lyse expired
expiration setting
The air inside the vacuum
Vacuum
filter
chamber is not extracted
error
within the given time.
Troubleshooting Your Analyzer
1. Stop using recorder and cool it for 10
2. If the problem remains, shut down the
1. If the problem remains, contact
1. Check if there is sufficient lyse left. If
2. If not, change a new container of lyse
1. Check if the diluent has expired. If so,
2. If not, reset the expiration date as
1. Check if the rinse has expired. If so,
2. If not, reset the expiration date as
1. Check if the lyse has expired. If so,
2. If not, reset the expiration date as
1. Enter "Service → Self–test →
2. If the test result is normal, the error will
3. If the problem remains, replace the
4. If the problem still remains after a new
11-8
minutes and then print again;
analyzer and restart it again, and then
print again;
Mindray customer service department.
so, access "Setup → Settings →
Reagents" and adjust the remaining
lyse volume as instructed by Chapter
5.3.1;
as instructed by Chapter 4.4.1.
change a new container of diluent as
instructed by Chapter 4.4.1;
instructed in Chapter 5.3.1.
change a new container of rinse as
instructed by Chapter 4.4.1;
instructed in Chapter 5.3.1.
change a new container of lyse as
instructed by Chapter 4.4.1;
instructed in Chapter 5.3.1.
Tubing"to test the filter as instructed
in Chapter 10.5.1;
be removed;
filter as instructed by Chapter 10.11;
filter has been installed, contact
Mindray customer service department