Mindray BC-2800 Operator's Manual page 109

Auto hematology analyzer
Hide thumbs Also See for BC-2800:
Table of Contents

Advertisement

Follow the previously introduced procedure to re-analyze the sample in question. The new
result will overwrite the old result while the sample information keeps unchanged.
Automatic saving of analysis results
This analyzer automatically saves a maximum of 10,000 sample results. When the maximum
number has been reached, the newest result will overwrite the oldest.
Parameter flags
If the analysis result is followed by an "H" or "L", it means the analysis result has
exceeded the upper or lower limit of the reference range.
If you see *** as opposed to the result, it means the result is either unreliable or out of the
operating range.
If the WBC result is less than 0.5 × 10
analysis and all the related parameter values will be non-numeric (***).
The result of the background check will not be flagged.
Histogram flags
The system will flag abnormal histograms.
Abnormal WBC histograms will be flagged by one of the markings: R1, R2, R3, R4 and
R
m.
R
: indicates abnormality on the left side of the lymphocyte hump and possible presence of
1
Operating Your Analyzer
Figure 6-11 "Re-count" screen
9
/L, this analyzer will not perform the differential
6-17