4.4 Installation Procedure

Dispose of reagents, waste, samples, consumables, etc. according to
government regulations.
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
Use the specified reagents.
After installing new reagents, keep the reagents still for a while before using
them.
Never use expired reagents.
To prevent contamination, tighten the container caps when the installation is
finished.
Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.

4.4.1 Connecting Regents

On the back of the instrument you will find a fluidic connection equipped with 4 plastic plugs.
These plugs are closed by protection caps to prevent dust and the leaking of liquids during
transportation. Take off these caps by unscrewing them and keep them in a safe place for
future transportation.
Connecting the diluent container
1. Take out the diluent pickup tube (the one with a green connector) from the accessory box
Installing Your Analyzer