6.8 Running Prediluted Samples
Press [MENU] and SELECT "Count" to enter the "Count" screen, as Figure 6-12 shows.
Figure 6-12 "Count" screen in prediluted mode
Press [MODE] to select prediluted mode (any mode preceded by Prediluted).
Select proper reference range as instructed in Chapter 5.3.6 before running
the samples. Otherwise, the obtained results may be erroneously flagged.
6.8.1 Entering Sample Information
You may enter sample information in any of the three modes, ID only, All info and Batch
edit, depending on which of the three configurations your analyzer has:
ID only
All Info
ID only and Batch edit.
ID only
To enter the sample ID of the next sample, you may
At the "Count" screen, use the bar-code scanner (if available) to scan the sample ID into the
analyzer; or
At the "Count" screen, press [F1] to enter the "ID" window and ENTER the sample ID.
Operating Your Analyzer
6-19