At either screen, you can press [←] or [→] to view the results (displayed below the parameter
box) of every point presented in the graph. The current cursor position is displayed to the
right of "No." and the time at which the sample was analyzed is displayed to the right of
"Time". You can also press [PgUp] or [PgDn] to jump forward or backward by 20 samples.
The trend is interpreted as follows:
The x-coordinate represents how many sample results have been selected. The
y-coordinate represents the analysis results of the displayed parameters.
For every parameter, the upper dash line of its trend represents the upper limit of the
expected range, 10% above the mean, of the analysis result. In case of WBC in Figure
7-45, the upper limit is "10.2".
Reviewing Sample Results
Figure 7-45 "Trend" screen (1)
Figure 7-46 "Trend" screen (2)