SELECT "Calibration → Fresh Blood" (Figure 9-10) to enter the "Fresh Blood" screen
(Figure 9-11).
Complete the fresh blood calibration as instructed below:
Determination of the reference values
Reference values may be determined using three or more normal blood samples prepared for
calibration from the reference method or from a reliably calibrated hematology analyzer.
Editing calibration settings
Selecting count mode
At "Fresh Blood" screen, press [F1] to select the "Whole Blood" or "Prediluted" mode.
Selecting sample
When the mode is selected, press [F3] to choose the fresh blood sample whose reference
values you want to set.
Editing the reference value
Press [F2] to enable the edit boxes in the "Mean" column, as Figure 9-12 shows.