Setting count time
If the WBC or RBC count time is inappropriately set, the system may give false alarms for
clogs or bubbles. When this happens, follow the steps given below to change the WBC or
RBC count time. Refer to the actual count time (see Chapter 10.5.1 for details) when editing
the count time. Follow the steps given below to set the count time.
1. Enter the administrator password as introduced in Chapter 5.2.1;
2. At the "Settings" screen, press [F1] to select the "Count" group;
3. SELECT "WBC" or "RBC", as Figure 5-31 shows;
4. ENTER the desired digits.
Exiting the "Count" group
Customizing the Analyzer Software
Figure 5-30 Selecting parameter unit
Figure 5-31 Setting count time
5-19