6.7 Running Whole Blood Samples

Press [MENU] and SELECT "Count" to enter the "Count" screen, as Figure 6-3 shows.
Press [MODE] to select whole blood mode (any mode preceded by Whole blood);
Select proper reference range as instructed in Chapter 5.3.6 before running
the samples. Otherwise, the obtained results may be erroneously flagged.
When you save the next sample information, run sample analysis;
Otherwise, the input sample information cannot be displayed when
restarting the analyzer.

6.7.1 Entering Sample Information

You may enter sample information in any of the three modes, ID only, All info and Batch
edit, depending on which of the three configurations your analyzer has:
ID only
All Info
ID only and Batch edit.
ID only mode
To enter the sample ID of the next sample, you may
Operating Your Analyzer
Figure 6-3 "Count" screen
6-9