Configuration and Use Manual
MMI-20025166
6.7.1 Options for Process Variable Fault Action
Display
ProLink III
Upscale
Upscale
Downscale
Downscale
Zero
Zero
Not-a-Number (NAN)
Not a Number
Flow to Zero
Flow to Zero
None (default)
None
Configuration and Use Manual
Configure advanced options for process measurement
Label
Field Communicator
Upscale
Downscale
IntZero-All 0
Not-a-Number
IntZero-Flow 0
None (default)
Description
•
Process variable values indicate that the
value is greater than the upper sensor limit.
•
Totalizers stop incrementing.
•
Process variable values indicate that the
value is lower than the lower sensor limit.
•
Totalizers stop incrementing.
•
Flow rate variables go to the value that
represents a flow rate of 0 (zero).
•
Density is reported as
.
0
•
Temperature is reported as
equivalent if other units are used (e.g.,
.
32°F
•
Drive gain is reported as measured.
•
Totalizers stop incrementing.
•
Process variables are reported as
.
NAN
•
Drive gain is reported as measured.
•
Modbus scaled integers are reported as
.
Max Int
•
Totalizers stop incrementing.
•
Flow rates are reported as
•
Other process variables are reported as
measured.
•
Totalizers stop incrementing.
•
All process variables are reported as
measured.
•
Totalizers increment if they are running.
March 2019
, or the
0°C
IEEE
.
0
115