Select The Active Concentration Matrix Using The Display; Set Up Concentration Measurement Using Prolink Iii - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20025166
Procedure
1. Choose Menu > Configuration > Process Measurement > Concentration Measurement > Configure
Matrix.
2. Select the matrix that you want to modify.
3. Set Extrapolation Limit to the point, in percent, at which an extrapolation alert will be posted.
4. Choose Menu > Configuration > Process Measurement > Concentration Measurement > Configure
Application > Extrapolation Alerts.
5. Enable or disable the high and low limit alerts for temperature and density as desired.
Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).

Select the active concentration matrix using the display

You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Menu > Configuration > Process Measurement > Concentration Measurement > Configure
Application.
2. Set Active Matrix to the matrix you want to use.

5.2.3 Set up concentration measurement using ProLink III

This section guides you through the tasks required to set up, configure, and implement concentration
measurement.
Enable the concentration measurement application using ProLink III
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
The concentration measurement application cannot be enabled at the same time as the API referral
application or the Advanced Phase Measurement application. They must be disabled first.
Procedure
1. Choose Device Tools > Configuration > Process Measurement > Flow and ensure that Volume Flow
Type is set to Liquid Volume.
2. Choose Device Tools > Configuration > Transmitter Options.
3. Set Concentration Measurement to Enabled and click Apply.
Configuration and Use Manual
Configure process measurement applications
March 2019
79