Set Up Energy Flow - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Measurement support
March 2019
Solve for A:
Solve for B:
Concentration slope (A): 1.00029
Concentration offset (B): 0.00551
14.9 Energy flow
Energy flow is the energy content of the process gas flowing through the pipe per units of time. The Model
5700 can compute the energy flow by polling for calorific value using a HART connection. The calculated
energy flow is available on the Energy Flow process variable.

14.9.1 Set up energy flow

Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART
communications.
If you plan to use the mA Input for the Calorific Value data, Channel D must be available for use and must
be wired to the external device providing Calorific Value.
Procedure
1. From the display, go to Menu > Configuration > Process Measurement > Energy Flow.
2. Select Units for Calorific Value and Energy Flow.
Set the Calorific Value units to match the Calorific Value units on the polled device (HART) or input
device (mA Input).
3. Set up polling or mA Input for the Calorific Value.
Option
Polling
mA Input
4. To set up polling:
224
50 = A × 49.98 + B
16 = A × 15.99 + B
50.00 − 16.00 = 34.00
49.98 − 15.99 = 39.99
34 = A × 33.99
A = 1.00029
50.00 = 1.00029 × 49.98 + B
50.00 = 49.99449 + B
B = 0.00551
Description
Go to
Step 4
Go to
Step 5
Micro Motion Model 5700 Transmitters with Configurable Outputs
Configuration and Use Manual
MMI-20025166