Drive Over Range; Electronics Failed; Event Active; Extreme Primary Purpose Variable - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configure device options and preferences
March 2019
Table 7-4: Core Low Power conditions
Name
[031] Low Power

Drive Over Range

Default severity: Maintenance
Severity configurable: Yes
Fault Timeout applicable: Yes
Table 7-5: Drive Over-Range conditions
Name
[102] Drive Overrange

Electronics Failed

Default severity: Failure
Severity configurable: No
Fault Timeout applicable: No
Table 7-6: Electronics Failed conditions
Name
[002] RAM Error (Core Processor)
[018] EEPROM Error (Transmitter)
[019] RAM Error (Transmitter)
[022] Configuration Database Corrupt (Core Processor)
[024] Program Corrupt (Core Processor)
Watchdog Error
Verification of mA Output 1 Failed

Event Active

Default severity: Out of Speculation
Severity configurable: Yes
Fault Timeout applicable: Yes
Table 7-7: Event Active conditions
Name
Enhanced Event [1 - 5] Active
Event [1 - 2] Active

Extreme Primary Purpose Variable

Default severity: Failure
130
Ignorable
No
Ignorable
Yes
Ignorable
No
No
No
No
No
No
No
Ignorable
Yes
Yes
Micro Motion Model 5700 Transmitters with Configurable Outputs
Configuration and Use Manual
MMI-20025166

Advertisement

Table of Contents
loading