Sd Card Full; Severe Two Phase Flow; Time Not Entered; Transmitter Software Update Failed - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
March 2019
Cause
The transmitter's SD card has failed.
Recommended actions
1. Make sure that the SD card is firmly seated in the socket.
2. Contact customer service.

17.13.61 SD Card Full

Alert
Data Loss Possible
Cause
The transmitter's SD card is 90% full.
Recommended actions
Contact customer service.

17.13.62 Severe Two Phase Flow

Alert
Process Aberration
Cause
The transmitter has detected severe two-phase flow.
Recommended actions
Check your process conditions against the values reported by the device.

17.13.63 Time Not Entered

Alert
Configuration Error
Cause
The system time has not been entered. The system time is required for diagnostic logs.
Recommended actions
Set the system time.

17.13.64 Transmitter Software Update Failed

Cause
The transmitter software update failed.
282
Micro Motion Model 5700 Transmitters with Configurable Outputs
Configuration and Use Manual
MMI-20025166

Advertisement