Configure process measurement applications
March 2019
Procedure
1. Choose Overview > Device Information > Licenses > Enable/Disable Applications and ensure that
Volume Flow Type is set to Liquid.
2. Choose Overview > Device Information > Licenses > Enable/Disable Applications.
3. Enable the concentration measurement application.
Load a concentration matrix from the transmitter's SD card using the Field
Communicator
If you have a concentration matrix on the transmitter's SD card, you can move it into one of the six slots on
your transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can
load up to six matrices into slots.
Prerequisites
You must have one or more concentration matrices loaded onto the transmitter's SD card.
You must know the derived variable that the matrix is designed to calculate.
Procedure
1. Choose Configure > Manual Setup > Measurements > Optional Setup > Conc Measurement > CM
Configuration and ensure that the setting of Derived Variable matches the derived variable used by
your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter's SD card. Verify the setting of Derived Variable before
continuing.
2. Choose Configure > Manual Setup > Measurements > Optional Setup > Conc Measurement > Load
Matrix File from IM.
3. Select the slot that you want to load to.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
4. Enter the name of the matrix file on the SD card, without the .matrix extension.
Example
If the matrix file name is test.matrix , enter test .
Postrequisites
Choose Configure > Manual Setup > Measurements > Optional Setup > Conc Measurement > CM
Configuration > Active Matrix and ensure that the selected slot contains the matrix that you loaded.
86
Micro Motion Model 5700 Transmitters with Configurable Outputs
Configuration and Use Manual
MMI-20025166