Configure advanced options for process measurement
March 2019
4. Select up to four totalizers or inventories to be logged.

6.7 Configure Process Variable Fault Action

Display
Menu > Configuration > Alert Setup > Output Fault Actions
Device Tools > Configuration > Fault Processing
ProLink III
Web browser
Configuration > Alert Setup > Output Fault Actions
Process Variable Fault Action specifies the values that will be reported via the display and digital
communications if the device encounters a fault condition. The values are also sent to the outputs for
processing against their configured fault actions.
Procedure
Set Process Variable Fault Action as desired.
Default: None
Restriction
If you set Process Variable Fault Action to NAN, you cannot set mA Output Fault Action or Frequency
Output Fault Action to None. If you try to do this, the transmitter will not accept the configuration.
Important
If you want the mA Output to continue reporting process data during fault conditions, you must set both
Process Variable Fault Action and mA Output Fault Action to None. If mA Output Fault Action is set to
None and Process Variable Fault Action is set to any other option, the mA Output will produce the signal
associated with the selection.
If you want the Frequency Output to continue reporting process data during fault conditions, you must set
both Process Variable Fault Action and Frequency Output Fault Action to None. If Frequency Output
Fault Action is set to None and Process Variable Fault Action is set to any other option, the Frequency
Output will produce the signal associated with the selection.
6.7.1 Options for Process Variable Fault Action
Display
ProLink III
Upscale
Upscale
Downscale
Downscale
92
Label
Web browser
Upscale
Downscale
Configuration and Use Manual
Description
Process variable values indicate that the
value is greater than the upper sensor limit.
Totalizers stop incrementing.
Process variable values indicate that the
value is lower than the lower sensor limit.
Totalizers stop incrementing.
Micro Motion Model 5700 with Ethernet Transmitters
MMI-20029769