[104] Calibration In Progress; [105] Two-Phase Flow; [110] Frequency Output X Saturated - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
March 2019
If it is not, correct the problem and reboot or power-cycle the transmitter.
If it is, replace the transmitter because the transmitter has an internal power issue.

17.13.26 [104] Calibration in Progress

Alert
Function Check in Progress
Cause
A calibration is running.
Recommended actions
Allow the test to complete.

17.13.27 [105] Two-Phase Flow

Alert
Process Aberration
Cause
The line density is outside the user-defined two-phase flow limits.
Recommended actions
1. Check for two-phase flow.
Refer to
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA
2. Verify that the transmitter is configured correctly for the connected sensor.

17.13.28 [110] Frequency Output X Saturated

Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the scaling of the Frequency Output.
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.
272
Micro Motion Model 5700 Transmitters with Configurable Outputs
Configuration and Use Manual
MMI-20025166
Output.