Emerson Micro Motion 5700 Configuration And Use Manual page 294

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
March 2019
g) Click Fix mA.
h) Read the mA current at the receiving device and compare it to the transmitter output.
The readings do not need to match exactly. If the values are slightly different, you can correct
the discrepancy by trimming the output.
i) Click UnFix mA.
2. Test the Frequency Output(s).
a) Choose Device Tools > Diagnostics > Testing and select the frequency output to test.
Note
If you have configured both Channel B and Channel D as Frequency Output 2, be sure to test
both loops.
b) Enter the frequency output value in Fix to.
c) Click Fix FO.
d) Read the frequency signal at the receiving device and compare it to the transmitter output.
e) Click UnFix FO.
3. Test the Discrete Output(s).
a) Choose Device Tools > Diagnostics > Testing > Discrete Output Test.
b) If your transmitter is configured for two or more discrete outputs, determine which discrete
output you want to test.
c) Set Fix to: to ON.
d) Verify the signal at the receiving device.
e) Set Fix to: to OFF.
f) Verify the signal at the receiving device.
g) Click UnFix.
4. Test the discrete input.
a) Set the remote input device to ON.
b) Choose Device Tools > Diagnostics > Testing > Discrete Input Test.
c) Verify the signal at the transmitter.
d) Set the remote input device to OFF.
e) Verify the signal at the transmitter.
5. Test the mA Input.
a) Set the remote input device to generate a known fixed current.
b) Choose Device Tools > Diagnostics > Testing > Read MA Input.
c) Return the remote input device to normal operation.
294
Micro Motion Model 5700 Transmitters with Configurable Outputs
Configuration and Use Manual
MMI-20025166