Emerson Micro Motion 5700 Configuration And Use Manual page 307

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20025166
Table 17-7: Velocity measurement problems and recommended actions
Problem
Possible causes
Inaccurate velocity
reading
Configuration and Use Manual
Wiring problem
Inappropriate measurement unit
Incorrect flow calibration factor
Incorrect density calibration factors
Incorrect grounding
Two-phase flow
Problem with receiving device
Incorrect sensor zero
(continued)
Recommended actions
Verify that the measurement units are
configured correctly for your application.
Zero the meter.
Check the grounding of all components.
Check for two-phase flow.
Verify the receiving device, and the wiring
between the transmitter and the receiving
device.
Replace the core processor or transmitter.