Table 17-7: Velocity measurement problems and recommended actions
Problem
Possible causes
Inaccurate velocity
•
reading
•
•
•
•
•
•
•
Configuration and Use Manual
Wiring problem
Inappropriate measurement unit
Incorrect flow calibration factor
Incorrect density calibration factors
Incorrect grounding
Two-phase flow
Problem with receiving device
Incorrect sensor zero
(continued)
Recommended actions
•
Verify that the measurement units are
configured correctly for your application.
•
Zero the meter.
•
Check the grounding of all components.
•
Check for two-phase flow.
•
Verify the receiving device, and the wiring
between the transmitter and the receiving
device.
•
Replace the core processor or transmitter.