4. Optional: For two-stage batching, assign one of the Discrete Outputs to the secondary valve.
You can also assign one of the Discrete Outputs to a pump when the batcher is configured for single
stage or two stage.
5.3.4 Configure AOC

Configure AOC using the display

AOC (Automatic Overshoot Compensation) adjusts the timing of the valve close command to minimize
overshoot. The valve close command is sent before the target is reached, to compensate for the time
required for the valve to close completely. AOC is applied to all presets and all batches.
There are three options for AOC:
Compensation Off
The transmitter sends the close command when the measured batch total reaches
the configured target.
Fixed
The transmitter sends the close command when the measured batch total equals the
Compensation
target minus the value configured for Fixed Overshoot Compensation. This value is
Value
configured in the current mass or volume unit, and applies to all presets.
AOC Algorithm
The transmitter compares the actual measured amount of each batch to the batch
target, and determines the adjustment by calibrating itself according to an internal
algorithm. You can choose to stop AOC calibration when you are satisfied with the
result, or you can set up rolling (continuous) AOC calibration.
Procedure
To use the Compensation Off option:
a) Choose Menu > Configuration > Process Measurement > Batcher > AOC.
b) Set Compensation Mode to Off.
To use the Fixed Compensation Value option:
a) Choose Menu > Configuration > Process Measurement > Batcher > AOC.
b) Set Compensation Mode to Fixed.
c) Set Fixed Value to the quantity to be subtracted from the batch target.
Important
Adjust the fixed value:
Whenever the process changes (e.g., flow rate, process fluid)
Whenever you change any configuration parameter that can affect flow measurement (e.g., damping)
To use the AOC Algorithm option:
a) Choose Menu > Configuration > Process Measurement > Batcher > AOC.
b) Set Compensation Mode to AOC.
c) Perform AOC calibration.
Configuration and Use Manual
Configure process measurement applications
March 2019
101