5.2.2 Set up concentration measurement using the display

This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. See the Micro Motion Enhanced Density Application
Manual for detailed information on building a matrix.
Enable the concentration measurement application using the display
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Procedure
1. Choose Menu > Configuration > Process Measurement.
2. Choose Flow Variables > Volume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the API referral application is displayed in the menu, choose API Referral and ensure that Enabled/
Disabled is set to Disabled.
The concentration measurement application and the API referral application cannot be enabled
simultaneously.
5. If the advanced phase measurement application is displayed in the menu, choose Advanced Phase
Measurement > Application Setup and ensure that Enabled/Disabled is set to Disabled.
The concentration measurement application and the advanced phase measurement application
cannot be enabled simultaneously.
6. Enable concentration measurement.
a) Choose Menu > Configuration > Process Measurement > Concentration Measurement.
b) Set Enabled/Disabled to Enabled.

Load a concentration matrix from a USB drive using the display

At least one concentration matrix must be loaded into one of the six slots on your transmitter. You can load
up to six matrices into slots. You can also copy matrices to the transmitter's SD card, and load them into slots
at a later time.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.