Load A Configuration File Using Prolink Iii - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Introduction to configuration and commissioning
March 2019
c) Select Backup or Replicate.
d) Select the file that you want to load.
e) Choose Yes or No when prompted to apply the settings.
Yes: The file is loaded to working memory and becomes active immediately.
No: The file is loaded to the transmitter's SD card but not to working memory. You can load it
from the SD card to working memory at a later time.

3.2.4 Load a configuration file using ProLink III

You can load a configuration file to the transmitter's working memory. You can load a backup file or a
replication file. Two PC file formats are supported: the Model 5700 format and the ProLink III format.
Backup files
Contain all parameters. They are used to restore the current device if required.
The .spare extension is used to identify backup files.
Replication files Contain all parameters except the device-specific parameters, e.g., calibration factors or
meter factors. They are used to replicate the transmitter configuration to other devices.
The .xfer extension is used to identify replication files.
Tip
You can use a saved configuration file to change the nature of the transmitter quickly. This might be
convenient if the transmitter is used for different applications or different process fluids.
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually
or by groups. Therefore, you can use this format for both backup and replication.
Procedure
To load a backup file or replication file from the transmitter's SD card:
a) Choose Device Tools > Configuration Transfer > Load Configuration.
b) Select On my 5700 Device Internal Memory and click Next.
c) Click Restore.
d) Set the file type.
To load a backup file, set the file type to Backup.
To load a replication file, set the file type to Transfer.
e) Select the file that you want to load and click Load.
The parameters are written to working memory, and the new settings become effectively immediately.
To load a backup file or replication file in Model 5700 format from the PC:
a) Choose Device Tools > Configuration Transfer > Load Configuration.
b) Select On my computer in 5700 device file format and click Next.
c) Click Restore.
28
Micro Motion Model 5700 Transmitters with Configurable Outputs
Configuration and Use Manual
MMI-20025166

Advertisement

Table of Contents
loading

Table of Contents