17.13.44 Batcher Not Configured

Alert
Configuration Error
Cause
One or more of the following has occurred:
The batcher application is disabled
No flow source has been configured for the batcher application
The batch target is 0
No discrete output is assigned to batch control.
Recommended actions
Verify the configuration of the batcher application and related parameters.

17.13.45 Batch Time Out

Alert
Process Aberration
Cause
The batch exceeded Maximum Batch Time, and ended before the target was reached.
Recommended actions
Verify the configuration of the batcher application and related parameters.

17.13.46 Clock Failure

Alert
Data Loss Possible
Cause
The transmitter's real-time clock is not incrementing.
Recommended actions
Contact customer service.

17.13.47 Core Processor Update Failed

Alert
Configuration Error