Emerson Micro Motion 5700 Configuration And Use Manual page 211

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20025166
Method
Description
mA Input
An external device sends
pressure data to the meter via
the mA Input.
Digital
A host writes pressure data to
communications
the meter at appropriate
intervals.
Postrequisites
Choose Menu > Service Tools > Service Data > View Process Variables and verify the external pressure value.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:
Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
For the mA Input:
Verify the wiring between the meter and the external device.
Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
Verify the Lower Range Value and Upper Range Value settings for the mA Input.
Perform an mA Input trim.
Adjust the mA Input damping value.
For digital communications:
Verify that the host has access to the required data.
Verify that the host is writing to the correct register in memory, using the correct data type.
Configuration and Use Manual
Setup
a. Choose Menu > Configuration > Process Measurement >
Pressure Compensation > External Pressure.
b. Set External Pressure to On.
c. Choose Menu > Configuration > Inputs/Outputs > Channel
D.
d. Set I/O Type to MAI.
e. Set Power appropriately.
f. Choose I/O Settings.
g. Set Assignment to External Pressure.
h. Set Lower Range Value and Upper Range Value
appropriately.
a. Choose Menu > Configuration > Process Measurement >
Pressure Compensation > External Pressure.
b. Set External Pressure to On.
c. Perform the necessary host programming and
communications setup to write pressure data to the
transmitter at appropriate intervals.
Measurement support
March 2019
211