Configure Process Measurement; Configure Sensor Flow Direction Arrow - Emerson Micro Motion 5700 Configuration And Use Manual

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Configuration and Use Manual
MMI-20025166

4 Configure process measurement

4.1 Configure Sensor Flow Direction Arrow

Display
Menu > Configuration > Process Measurement > Flow Variables > Flow Direction
ProLink III
Device Tools > Configuration > Process Measurement > Flow > Sensor Direction
Field Communicator
Configure > Manual Setup > Measurements > Flow > Sensor Direction
Sensor Flow Direction Arrow is used to accommodate installations in which the Flow arrow on the sensor
does not match the majority of the process flow. This typically happens when the sensor is accidentally
installed backwards.
Sensor Flow Direction Arrow interacts with mA Output Direction, Frequency Output Direction, and
Totalizer Direction to control how flow is reported by the outputs and accumulated by the totalizers and
inventories.
Sensor Flow Direction Arrow also affects how flow is reported on the transmitter display and via digital
communications. This includes ProLink III, the Field Communicator, and all other user interfaces.
Figure 4-1: Flow arrow on sensor
A. Flow arrow
B. Actual flow direction
Procedure
Set Sensor Flow Direction Arrow as appropriate.
Option
Description
With Arrow
The majority of flow through the sensor matches the Flow arrow on the sensor. Actual
forward flow is processed as forward flow.
Configuration and Use Manual
Configure process measurement
March 2019
31