17.13.20 [034] Smart Meter Verification Failed

Alert
Function Check Failed or Smart Meter Verification Aborted
Cause
The current SMV value is statistically different than the factory baseline value.
Recommended actions
Minimize process instability and repeat the test.

17.13.21 [035] Smart Meter Verification Aborted

Alert
Function Check Failed or Smart Meter Verification Aborted
Cause
The SMV test did not complete, possibly because it was manually aborted or because process conditions were
too unstable.
Recommended actions
1. Minimize process instability and repeat the test.
2. Check the SMV abort code and take appropriate steps.
3. Contact customer service.

17.13.22 [100] mA Output 1 Saturated

Alert
Output Saturated
Cause
The calculated output value is outside the range of the output.
Recommended actions
1. Check the settings of Upper Range Value and Lower Range Value.
Refer to
Configure Lower Range Value (LRV) and Upper Range Value (URV) for the mA
2. Check your process conditions against the values reported by the device.
3. Ensure that both devices are using the same measurement unit.
4. Purge the sensor tubes.