Emerson Micro Motion 5700 Configuration And Use Manual page 296

Transmitters with configurable outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Troubleshooting
March 2019
h) Choose End.
2. Test the Frequency Output(s).
a) Choose Service Tools > Simulate > Simulate Outputs and select the frequency output to test.
Channel B = Frequency Output 2
Channel C = Frequency Output 1
Channel D = Frequency Output 2
Note
If you have configured both Channel B and Channel D as Frequency Output 2, be sure to test
both loops.
b) Select the frequency output level.
c) Press OK.
d) Choose End.
3. Test the Discrete Output(s).
a) Choose Service Tools > Simulate > Simulate Outputs and select the discrete output to test.
Channel B = Discrete Output 1
Channel C = Discrete Output 2
Channel D = Discrete Output 3
b) Choose Off.
c) Verify the signal at the receiving device.
d) Press OK.
e) Choose On.
f) Verify the signal at the receiving device.
g) Press OK.
h) Choose End.
4. Test the Discrete Input.
a) Set the remote input device to ON.
b) Choose Service Tools > Variables > Inputs/Outputsand read the state of the discrete input.
Channel C = Discrete Input 1
Channel D = Discrete Input 2
c) Set the remote input device to OFF.
d) Choose Service Tools > Variables > Inputs/Outputsand read the state of the discrete input.
Channel C = Discrete Input 1
296