Preparation
Whenever the breathing hoses or the breathing
gas humidifier have been changed
Check the breathing circuit, see "Performing the
breathing circuit check" on page 66.
Additional information
For the order numbers of the breathing circuits and
the hinged arm, see the list of accessories.
Installing a neonatal flow sensor
The following neonatal flow sensors are available:
– Neonatal flow sensor ISO 15 (8411130)
– Neonatal flow sensor Y-piece (8410185)
Installing an ISO 15 neonatal flow sensor
(8411130)
1 Insert the neonatal flow sensor (A) into the pa-
tient connector of the Y-piece (B).
2 Connect plug (C) of the flow sensor cable to the
flow sensor.
48
C
B
A
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Installing a neonatal flow sensor Y-piece
(8410185)
1 Connect Y-piece with integrated neonatal flow
sensor (D) to the breathing hoses.
2 Connect plug (E) of the flow sensor cable to the
flow sensor.
Further procedure for both neonatal flow
sensors
3 Position patient connector of the Y-piece
about 45° downwards to prevent condensation
from forming on the neonatal flow sensor.
4 Route the cables along the breathing hoses to
the device.
G
5 Insert the connector (F) of the flow sensor cable
into the socket (G) at the rear of
Babylog VN500.
E
D
F