Operation
WARNING
Fire hazard!
Before medication nebulization, remove the
neonatal flow sensor from the Y-piece.
The wires of the neonatal flow sensor are hot.
If the flow sensor is left in the breathing circuit
during nebulization, medication aerosol de-
posits may build up and impair flow measure-
ment. In the worst case, these deposits could
catch fire!
Disconnecting the neonatal flow sensor cable
is not sufficient to prevent this.
The minute volume is not monitored without
the neonatal flow sensor and apnea monitor-
ing is limited. Use additional monitoring.
When using the Y-piece with integrated neonatal
flow sensor (8410185):
F
1 Disconnect the cable (F) from the neonatal flow
sensor (G).
96
2 Remove the insert (H).
3 Insert the sealing plug (I) (8411024). The seal-
When using the ISO 15 neonatal flow sensor
(8411130):
G
1 Remove the flow sensor (J) (housing and insert)
2 Connect the tube catheter cone (K) to the
After removing the neonatal flow sensor:
Babylog VN500 starts nebulization. The
bol and the remaining nebulization time is dis-
played in the screen header bar.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
H
I
ing plug is a component of the medication neb-
ulizer.
J
from the Y-piece.
Y-piece.
Replace or clean the neonatal flow sensor if
there is visible soiling. See "Dismantling neona-
tal flow sensor" on page 194.
Touch the Done (E) button.
K
sym-