Assembling The Inspiratory Unit; Assembling Accessories - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Assembling the inspiratory unit

1 Make sure the inspiratory unit and diaphragm
are completely dry, otherwise this may impair
proper functioning.
2 Insert the adapter (A) of the diaphragm into the
opening of the fitting (B). The adapter must be
able to slightly move up and down in the open-
ing.
3 Position the diaphragm in such a way that it is in
the recesses (C) of the fitting.
4 Fit the diaphragm onto the edge of the fitting
(D).
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
A
C
B
C
D
Cleaning, Disinfection and Sterilization
Inserting the inspiratory unit into
Babylog VN500
1 Insert the inspiratory unit (E) into the recesses
of the fitting and push as far as it will go into the
fitting.
2 Turn the inspiratory unit in clockwise direction
until the lock clicks into place.
3 Check whether the inspiratory unit is properly
engaged.

Assembling accessories

Assemble the medication nebulizer and breathing
gas humidifier in accordance with the correspond-
ing Instructions for Use.
Connecting the medication nebulizer to the
breathing circuit, see page 94.
Preparing the breathing gas humidifier,
see page 46.
E
205