Dräger Babylog VN500 Instructions For Use Manual page 169

Infinity acute care system
Table of Contents

Advertisement

Alarm
Alarm
priority
!!!
181
Apnea
!!
230
Apnea Ventilation
!
020
Application already in-
stalled
!
020
Application transfer failed
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Cause
The patient has stopped
breathing.
Obstruction.
Flow sensor is not calibrated
or faulty.
Due to detected apnea, the
ventilation unit has automat-
ically switched to Apnea
Ventilation.
Application is already in-
stalled.
Invalid application.
Application installation
failed.
Alarm – Cause – Remedy
Remedy
Check patient condition.
Apply controlled ventilation if
necessary.
Check patient condition.
Check breathing circuit.
Check tube or mask.
Calibrate flow sensor and
replace it if necessary.
Check patient condition.
Check tube or mask.
Check ventilation settings
and patient condition. Re-
turn to the original ventila-
tion mode by touching the
"Apn. Vent. reset" button
and confirm with rotary
knob.
Acknowledge message by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
Acknowledge message by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
Call DrägerService.
Acknowledge message by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
Call DrägerService.