PC-AC

Pressure Control-Assist Control
Assist-controlled, pressure-controlled ventilation
allowing spontaneous breathing during the entire
respiratory cycle and back-up respiratory rate
Paw
PEEP
Flow
Pressure-controlled ventilation
The upper pressure level is determined by Pinsp.
The duration of the mandatory breaths is deter-
mined by Ti. As in all pressure-controlled ventilation
modes, the tidal volume supplied depends on the
difference in pressure "Pinsp – PEEP", the lung
mechanics (resistance and compliance) and the
patient's respiratory drive. The pressure rise from
the lower pressure level PEEP to the upper pres-
sure level Pinsp is determined by the Slope or In-
sp. flow setting. The start setting can be configured
on the page System setup > Ventilation > Start
settings > Pressures, O
Assisted-controlled ventilation
Every inspiratory effort of the patient on PEEP level
triggers a synchronized mandatory breath. Thus,
the time and number of mandatory breaths are de-
termined by the patient. The trigger window covers
the expiratory time minus a refractory period for the
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
slow
fast
pressure rise
pressure rise
Pinsp
Ti
1
RR
without spontaneous
breathing
, I:E.
2
Trigger window for
inspiratory synchronization
with spontaneous breathing
previous expiration. The expiratory time is deter-
mined by the respiratory rate RR and the inspirato-
ry time Ti. A non-synchronized mandatory breath is
triggered at the latest at the end of the expiratory
time (backup respiratory rate).
The minimal number of mandatory breaths is deter-
mined by the respiratory rate RR.
Description
t
t