Reprocess housing of the neonatal flow sensor
in accordance with the reprocessing list,
see page 202.
Dismantling neonatal flow sensor Y-piece
(8410185)
1 Disconnect the flow sensor cable (A) from the
neonatal flow sensor.
2 Unplug the Y-piece with integrated neonatal
flow sensor (B) from the breathing hoses.
Reprocessing neonatal flow sensor Y-piece
(8410185)
Observe related instructions for use.

Dismantling the pneumatic medication

nebulizer

E
D
1 Remove the nebulizer hose (A) from the medi-
cation nebulizer (B) and from the nebulizer port
on the device.
2 Remove the medication nebulizer (B) from the
breathing circuit.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
A
B
C
A
B
Cleaning, Disinfection and Sterilization
3 Remove the catheter connector (C) from the in-
let port.
4 Pull the adapter (D) out of the outlet port.
5 Remove the corrugated hose (E) from the
adapter (D).
6 Dismantle the medication nebulizer in accor-
dance with the corresponding Instructions for
Use.
Reprocessing the medication nebulizer and
parts for adapting
Reprocess the individual parts of the medica-
tion nebulizer in accordance with the corre-
sponding Instructions for Use.
Reprocess the parts for adapting in accordance
with the reprocessing list, see page 202.
CO
sensor
2
Dismantling the CO
2
1 Remove the CO
sensor plug from the socket.
2
A
C
B
2 Remove the CO
sensor (A) from the cuvette.
2
3 Remove the cuvette (B) of the CO
the Y-piece.
4 Remove the catheter cone (C) from the cuvette.
sensor
sensor from
2