Maintenance

Maintenance intervals for Babylog VN500

CAUTION
Disinfect and clean the device and its components
before all maintenance work, also before return-
ing device for repairs!
The following table provides an overview of the
maintenance intervals for Babylog VN500.
Ambient air filter (Babylog VN500)
Ambient air filter (PS500)
Gas supply unit GS500
Diaphragm of the expiratory valve
Internal battery (Babylog VN500)
Internal batteries (PS500)
O
filter (in the O
gas inlet)
2
2
Air filter (in the Air gas inlet)
Device Inspection and Servicing
Safety inspections
208
Clean or exchange after 4 weeks, see "Exchanging the
ambient air filter" on page 210.
Exchange after 1 year.
Dispose of with normal domestic waste.
Must be serviced in conjunction with annual inspections.
Air filter of blower unit must be replaced annually by trained
service personnel.
Filter mat must be replaced annually by trained service
personnel.
Exchange after 1 year, see "Exchanging the diaphragm of
the expiratory valve" on page 210.
Dispose of with normal domestic waste.
Must be serviced in conjunction with annual inspections.
To be exchanged at the latest every 2 years by trained
service personnel.
Check capacity every year; have battery exchanged by
trained service personnel if necessary.
Must be serviced in conjunction with annual inspections.
To be exchanged at the latest every 2 years by trained
service personnel.
Check capacity every year; have batteries replaced by
trained service personnel if necessary.
To be exchanged every 6 years by trained service
personnel.
To be exchanged every 2 years by trained service
personnel.
Every year by trained service personnel.
Every year by trained service personnel. Scope see "Safety
inspections" on page 209.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n