Intensive care ventilator for neonates (174 pages)
Summary of Contents for Dräger Oxylog 3000 plus
Page 1
Instructions for use Oxylog 3000 plus Emergency and Transport Ventilator WARNING To properly use this medical device, Software 1.n read and comply with these instructions for use.
Page 2
NOTE The reproductions of screen content in the A NOTE provides additional information intended instructions for use can differ from the content to avoid inconvenience during operation. actually shown on the screen. Instructions for use Oxylog 3000 plus SW 1.n...
Page 3
Users are intended operators as defined on page 14 hereof for the use of the medical device in accordance with its intended use. Abbreviations and symbols For explanations refer to "Abbreviations" on page 23 and "Symbols" on page 24. Instructions for use Oxylog 3000 plus SW 1.n...
Page 4
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 5
. Displaying configuration and information . . . Attaching the Oxylog 3000 plus to standard rail Customer Service Mode ....systems ......
Page 6
Disassembly ......122 Oxylog 3000 plus SW 1.n ....173 Information on reprocessing.
Page 7
Product-specific safety information ..Installing accessories ....Only one copy of instructions for use included Instructions for use Oxylog 3000 plus SW 1.n...
Page 8
Only use the medical device together with authorized accessories listed in the current list of accessories. Instructions for use Oxylog 3000 plus SW 1.n...
Page 9
Instructions for use Oxylog 3000 plus SW 1.n...
Page 10
Medical electrical equipment Part 1-2: General requirements for safety Collateral standard: Electromagnetic compatibility; Requirements and tests – IEC 60601-1-4 Medical electrical equipment Part 1-4: General requirements for safety Collateral standard: Programmable electrical medical systems Instructions for use Oxylog 3000 plus SW 1.n...
Page 11
The user must ensure that the device combination meets the applicable standards. Strictly observe instructions for use and assembly instructions of all connected devices. Instructions for use Oxylog 3000 plus SW 1.n...
Page 12
The user must not solely rely on the built-in CAUTION monitoring of the ventilator and must always Installations on the Oxylog 3000 plus must be assume full responsibility for proper done in accordance with these instructions for ventilation and patient safety in all situations.
Page 13
Indications/Contraindications ... Environment of use ....Instructions for use Oxylog 3000 plus SW 1.n...
Page 14
Intended use The Oxylog 3000 plus is a time-cycled, volume- controlled and pressure-controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards. Intended user: the device is intended for use by and under the supervision of trained healthcare professionals, e.g.
Page 15
– The patient must be immediately transferred to a breathable atmosphere in order to prevent inhalation of infectious air when spontaneous breathing resumes. Instructions for use Oxylog 3000 plus SW 1.n...
Page 16
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
100 % O application (factory set) L Insp. Hold key for initiating a manual inspiration or for extending the current inspiratory time M Start/Standby key Instructions for use Oxylog 3000 plus SW 1.n...
F Connector for the DC power supply Tilting the device to a vertical position may lead to airway pressure oscillation. G Connector for the CO sensor H Connector for data communication cable Instructions for use Oxylog 3000 plus SW 1.n...
A Angled connector B Flow sensor B Flow sensor C Breathing valve C Breathing valve D Breathing hose D Breathing hose E Flow and pressure measuring lines E Flow and pressure measuring lines Instructions for use Oxylog 3000 plus SW 1.n...
System overview Disposable breathing circuit for pediatric patients A Angled connector B Flow sensor C Breathing valve D Breathing hose E Flow and pressure measuring lines Instructions for use Oxylog 3000 plus SW 1.n...
– NIV: in the ventilation modes: Spn-CPAP (/PS), PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF, and VC-SIMV / AF. Special functions: – Inspiration hold, – O inhalation (optional), with an inhalation mask – 100 % O (optional) Instructions for use Oxylog 3000 plus SW 1.n...
Time before apnea is recognized IPX4 Ingress Protection level 4 Expiratory time MEDIBUS.X Dräger communications protocol for medical devices with uniform Inspiratory time data definition for all devices Tplat% Plateau time in % of inspiratory time Instructions for use Oxylog 3000 plus SW 1.n...
Key for silencing acoustic alarm documents signals for 2 minutes Protection class type BF (Body Key for acknowledging alarm Alarm Floating), defibrillator-proof messages Reset Charge status of the internal battery Rotary knob Instructions for use Oxylog 3000 plus SW 1.n...
Page 25
Warning! Strictly follow these instructions for use Adult Adult For dry locations only Slope (steep, medium, flat) Warning, dangerous voltage! Quantity Temperature limitations Do not use if package is damaged Latex free Ambient pressure Instructions for use Oxylog 3000 plus SW 1.n...
Page 26
System overview Symbol Explanation Relative humidity Part number Lot number Recycling in accordance with the regulations for lithium-ion batteries. UN Regulation No. 10, revision 3 (EMC) 10 R - 03 XXXX Instructions for use Oxylog 3000 plus SW 1.n...
Page 27
Additional function keys ....On-screen window structure ... . Instructions for use Oxylog 3000 plus SW 1.n...
Operating concept Switching on and off Oxylog 3000 plus Switching on To switch the device ON, briefly press the key (A). Switching off Refer to the "Shutdown" on page 81. Instructions for use Oxylog 3000 plus SW 1.n...
The active ventilation mode is displayed in the – Respiratory rate RR [/min], upper left corner of the display (C). – Maximum inspiratory pressure Pmax [mbar], – Inspiratory O concentration FiO [%]. Instructions for use Oxylog 3000 plus SW 1.n...
(ventilation screen) in the settings window and for switching between screen pages E Alarms key for displaying the alarm settings in the alarms window and for switching between screen pages Instructions for use Oxylog 3000 plus SW 1.n...
C Insp. Hold key for initiating a manual inspiration or for extending the current inspiratory time. inhalation key for O inhalation or 100 % O key for 100 % O application (factory set). Instructions for use Oxylog 3000 plus SW 1.n...
E Window for settings or alarms E Trigger indicator. F Information window. For information on the The last page shows an overview of all measured content, refer to "Messages in the information values. field" on page 118. Instructions for use Oxylog 3000 plus SW 1.n...
Page 33
– RRapn and VTapn – Slope – Tapn – Tplat – Trigger – Psupp B Number of pages (1st page of 3 available pages) Press Settings key. The pages are displayed consecutively. Instructions for use Oxylog 3000 plus SW 1.n...
Page 34
Paw. Curves Curves Press Curves key multiple times. Enlarged view of flow curve Curve window showing the flow curve. Enlarged view of CO curve Curve window showing the CO curve. Instructions for use Oxylog 3000 plus SW 1.n...
Page 35
......Attaching the Oxylog 3000 plus to standard rail systems....Instructions for use Oxylog 3000 plus SW 1.n...
Page 36
Do not use an adult breathing hose for tidal volumes below 100 mL. CAUTION Do not use electrically conductive hoses. This may endanger the user and may cause damage to the device during defibrillation. Risk of electric shock. Instructions for use Oxylog 3000 plus SW 1.n...
3 Remove the battery (A) by pulling the tab. Checking the charge status of the battery Press the button on the rechargeable battery. The charge status is indicated by four LEDs. Instructions for use Oxylog 3000 plus SW 1.n...
It is recommended to have a fully charged spare battery available each time when using the Oxylog 3000 plus. Always position the device so that the external power connector can be easily disconnected from the ventilator. Instructions for use Oxylog 3000 plus SW 1.n...
1 Plug the large connector (A) of the DC/DC protected by a 10 to 16 A DC fuse. converter into the on-board supply. Outside this range the Oxylog 3000 plus cannot use the DC input power. 2 Plug the small connector (B) into the DC connector (C) of the Oxylog 3000 plus.
1 Connect the mains plug (A) to the mains outlet. 2 Connect the DC connector (B) to the DC connector (C) of the Oxylog 3000 plus. 3 When the Oxylog 3000 plus is correctly connected to an external power supply, the indicator (D) lights up.
O concentration below 25 %, to prevent risk of fire. WARNING No smoking or open flames. enhances combustion of other substances and can intensify fires. Instructions for use Oxylog 3000 plus SW 1.n...
Do not install flow control valves or flowmeters in the gas supply to the Oxylog 3000 plus. WARNING Always check the O pressure of cylinder before use, to prevent insufficient oxygen supply during use. Instructions for use Oxylog 3000 plus SW 1.n...
(C). Note the correct alignment of the parts by the groove in the flow sensor (D) and the notch on the breathing valve (C). 4 Push the angled connector (E) onto the flow sensor (D). Instructions for use Oxylog 3000 plus SW 1.n...
2 Connect the flow measuring lines (B) to the nozzles of the flow sensor. When connecting the flow measuring lines, pay attention to the differing diameters of the hoses and nozzles and connect them on the correct side. Instructions for use Oxylog 3000 plus SW 1.n...
Oxylog 3000 plus. When connecting a breathing hose, check that the hose setting in the settings window corresponds to the connected hose. NOTE Disposable hoses are shipped clean but non- sterile. Instructions for use Oxylog 3000 plus SW 1.n...
3 Connect the breathing hose (C) to the gas outlet on the Oxylog 3000 plus. When connecting a breathing hose, check that the hose setting in the settings window corresponds to the connected hose. Instructions for use Oxylog 3000 plus SW 1.n...
HME. Refer to the "Setting the HME correction" on page 79. Connect the bacterial filter or HME to the angled Disposable pediatric hose connector as follows. Adult reusable hose Instructions for use Oxylog 3000 plus SW 1.n...
(D). The CO sensor cable can be extended with a maximum of one extension cable. Refer to the "List of accessories" on page 165. Instructions for use Oxylog 3000 plus SW 1.n...
Page 49
Be careful when placing the ventilator on the rail or bed rim. CAUTION The Oxylog 3000 plus is only held by its own weight when hung on a bar or rail. The Oxylog 3000 plus must be secured additionally when being transported, otherwise vibrations may cause accidental dislodgement.
Page 50
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 51
Zero calibration before ventilation ..filter check before ventilation ..Preparation for use after system check, zero calibration and CO2 filter check . Instructions for use Oxylog 3000 plus SW 1.n...
When an external supply is available: Oxylog 3000 plus The remaining capacity of the battery is indicated by the Oxylog 3000 plus in 25 % increments in the lower right section of the information window when power is ON. As an example, in the above screen the battery is 75 % charged.
The amount of gas from the high-pressure supply, which is currently being consumed, is indicated by the Oxylog 3000 plus in the lower left section of the information window in L/min. This display is overwritten when a higher priority message is activated.
Make no changes afterwards, otherwise there is the risk of – At least every six months. inaccurate flow measurement and/or The Oxylog 3000 plus interrupts the device check if insufficient ventilation. a fault is detected. WARNING The relevant fault is indicated on the screen.
2 Press the rotary knob (B) to confirm, before the NOTE progress bar is complete. The start-up screen The Oxylog 3000 plus determines MVe and VTe appears: values assuming BTPS conditions, see page 141. Consequently, the measured values of a patient...
During the automatic test sequence, the measuring lines from the ventilator. Oxylog 3000 plus checks the flow, pressure 3 Check if an applicable alarm occurs. levels and alarm signals. Corresponding sounds are heard. Instructions for use Oxylog 3000 plus SW 1.n...
If the device check is not completed successfully: 1 Refer to "Error messages during the device check" on page 120 of the "Problem solving" section. 2 Call DrägerService. Instructions for use Oxylog 3000 plus SW 1.n...
If the check was not successful: If zero calibration was not successful: The Oxylog 3000 plus displays the alarm Filter check failed. The measured CO value is outside The Oxylog 3000 plus displays the alarm message the permissible tolerance range.
Preparation for use after system check, CO zero calibration and CO2 filter check 1 Assemble the Oxylog 3000 plus for operation. Refer to the "Assembly" on page 35. 2 Connect to the power supply and gas supply. Refer to "Internal rechargeable battery" on page 37 and "Connecting the gas supply"...
Page 61
Calibration......Screen brightness ....Instructions for use Oxylog 3000 plus SW 1.n...
Operation Starting operation The actual screen display may differ in appearance The Oxylog 3000 plus performs a selftest. The or configuration. selftest will be completed in approximately 6 seconds. WARNING During the selftest the start-up screen is briefly Only use a ventilator that has been cleaned, displayed.
Page 63
Operation Screen with default settings The start-up settings can be configured in Customer Service Mode. Refer to the "Setting the start-up settings" on page 98. Instructions for use Oxylog 3000 plus SW 1.n...
I:E will be shown in the information window when Ti is selected, and vice versa. This configuration will apply to all ventilation modes. Refer to the "Customer Service Mode" on page 97. Instructions for use Oxylog 3000 plus SW 1.n...
– Respiratory rate RR – Plateau time Tplat as a percentage of the (minimum possible respiratory rate: 5 /min) inspiratory time – Maximum airway pressure Pmax – AutoFlow (optional) – O concentration FiO Instructions for use Oxylog 3000 plus SW 1.n...
2 Select the line Trigger on the display and then set and confirm the value with the rotary knob. Low value = high sensitivity. The ventilation mode AC is shown on the display. Instructions for use Oxylog 3000 plus SW 1.n...
2 Press the rotary knob to confirm. compressions. The last effective trigger value is adopted by the The Oxylog 3000 plus will try to limit the airway ventilator when changing from VC-AC to PC-BIPAP pressure Paw to the set Pmax, without ending the or SpnCPAP.
– Plateau time Tplat as a percentage of the – Respiratory rate RR inspiratory time (minimum possible respiratory rate: 2 /min). – AutoFlow (optional) – Maximum airway pressure Pmax – O concentration FiO Instructions for use Oxylog 3000 plus SW 1.n...
±8 mmHg CO When AutoFlow is switched on, the setting Tplat is no longer valid, and Slope must be set. For more information on AutoFlow, refer to "AutoFlow" on page 160. Instructions for use Oxylog 3000 plus SW 1.n...
Refer to the description on page 159 for details. Refer to "Non-invasive ventilation (NIV)" on Set the ventilation pattern with the controls below page 75. the display: – Respiratory rate RR – Maximum airway pressure Pmax Instructions for use Oxylog 3000 plus SW 1.n...
Therefore, the correct cuvette type must be selected in the settings menu. Otherwise the zero point will be shifted by as much as ±8 mmHg CO Instructions for use Oxylog 3000 plus SW 1.n...
Therefore, the correct cuvette type must be selected in the settings menu. Otherwise the zero point will be shifted by as much as ±8 mmHg CO Instructions for use Oxylog 3000 plus SW 1.n...
Pmax. The ventilation time ratio I:E = 1:1.5 and the plateau time Tplat = 0 are preset during apnea 2 Set RRapn and VTapn. ventilation. 3 Set Pmax to set the maximum permissible airway pressure during apnea ventilation. Instructions for use Oxylog 3000 plus SW 1.n...
Spn-CPAP: – Trigger sensitivity Successful patient triggering is indicated by an asterisk (*) on the left side of the curve window. – Pressure rise time Slope (effective for Pressure Support Psupp) Instructions for use Oxylog 3000 plus SW 1.n...
Spn-CPAP (/PS), 2 Select NIV on and confirm. PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and VC-SIMV / AF. The Oxylog 3000 plus automatically – The suffix NIV appears in the ventilation mode adjusts to the requirements of non-invasive field.
0 mbar. The !!! Paw high alarm is active. NOTE The options 100 % O and O inhalation are Press the Insp. Hold (A) key for the desired mutually exclusive. inspiratory time. Instructions for use Oxylog 3000 plus SW 1.n...
Page 77
2 Press and hold key O inhalation (A) for approx. 3 seconds. inhalation is performed with the previously effective setting. 3 Set and confirm the required O flow via the rotary knob (C). Display (example): Instructions for use Oxylog 3000 plus SW 1.n...
It is not measured by an internal O sensor. 82 (%) If the Oxylog 3000 plus is unable to reach the set concentration, the user is prompted to correct the setting by the message Check settings: FiO Flow ...
The temperature and moisture influence of the HME (heat and moisture exchanger) have an effect on the flow measurement. The Oxylog 3000 plus can compensate for the presence of an HME. When using HME, select, set and confirm HME –...
(power save mode). Alarm volume The alarm volume can be set on the last page of the alarm menu, from level 1/4 to 4/4. Instructions for use Oxylog 3000 plus SW 1.n...
3 With gas supply from an O cylinder: Close the cylinder valve. With gas supply from the pipeline gas supply system: Disconnect the high-pressure connection from the source. Instructions for use Oxylog 3000 plus SW 1.n...
Page 82
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 83
Alarm indication ..... . Setting alarm limits ....Instructions for use Oxylog 3000 plus SW 1.n...
Page 84
Alarm priorities The actual screen display may differ in appearance or configuration. The Oxylog 3000 plus assigns priorities to alarms. The alarm message is highlighted with the Oxylog 3000 plus corresponding number of exclamation marks. Also, different acoustic alarm signals are generated for the various alarm priorities.
Page 85
Alarms Warnings are preceded by three exclamation The Oxylog 3000 plus generates a 3-tone marks and displayed in inverted form (C). acoustic alarm signal that is repeated about every 20 seconds. Example: !!! Apnea ventilation The Oxylog 3000 plus generates a 5-tone...
Page 86
NOTE To be notified of new acoustic alarm signals, the 2 minutes alarm silencing must be reset. To cancel the silencing before the 2 minutes elapse: Press the key (A) again. Instructions for use Oxylog 3000 plus SW 1.n...
Page 87
67. Pmax appears in the pressure curve as a dashed line. When this dashed line is reached, the Oxylog 3000 plus issues a !!! Paw high alarm. 2 Select and activate the low alarm limit the upper alarm limit...
Page 88
0.5 L/min the limit remains value –1.3 +0.7 unchanged. Measured value +30 % or +2 L/min, whichever is smaller. RRsp Measured value +5/min with a minimum of 10/min. etCO Based on the current value Instructions for use Oxylog 3000 plus SW 1.n...
Page 89
Serial port ......Requirements for the electrical properties of connected devices and networks ..Instructions for use Oxylog 3000 plus SW 1.n...
The values are shown in pairs; the pairs of values can be configured as required. Refer to the "Customer Service Mode" on page 97. Instructions for use Oxylog 3000 plus SW 1.n...
– Alarm window: etCO high and etCO NOTE alarms If a wrong cuvette type is selected, the Oxylog 3000 plus displays the alarm !!! Check cuvette type. For more information on curves and measured values: refer to "Enlarged view of CO curve" on page 34.
Monitoring Performing zero calibration during If zero calibration was not successful: ventilation The Oxylog 3000 plus displays the alarm !!! CO Zero calib. failed. The zero calibration is performed with a clean CO Redo the zero calibration. sensor that has been removed from the cuvette!
Pass. 6 Attach the CO sensor (A) back to the cuvette. If the check was not successful: The Oxylog 3000 plus displays the alarm !!! CO Filter check failed. The measured CO value is outside the permissible tolerance range.
– IEC 60950-1: Ungrounded SELV circuits MEDIBUS.X is a software protocol for the transfer – IEC 60601-1 (from 2nd edition): Touchable of data between the Oxylog 3000 plus and an secondary circuits external medical or non-medical device (e.g., patient monitors or computers for data management systems).
Page 95
Exiting Customer Service Mode ... 107 Customer service manual ....108 Instructions for use Oxylog 3000 plus SW 1.n...
– Identification No. (Device ID) – Total hours of operation (Working hours) – Hours of operation since the last inspection and maintenance (Hours since service time) – Battery type and battery capacity. Instructions for use Oxylog 3000 plus SW 1.n...
Display the maintenance schedule and contact informa- contact information. tion for maintenance personnel. sensor. Check or calibrate the CO sensor. Acoustic alarm configuration Set the minimum alarm volume and alarm tone type. Instructions for use Oxylog 3000 plus SW 1.n...
VTapn (ped.) 100 mL Alarm volume 1/4 to 4/4 (VTapn (ped.)) (Alarm volume) RRapn (RRapn) 12 /min etCO -high alarm 1 to 100 mmHg RRapn (ped.) 27 /min (etCO -high alarm) (RRapn (ped.)) Instructions for use Oxylog 3000 plus SW 1.n...
Page 99
The start-up settings for the parameters are displayed on the screen when the ventilator is switched ON. The settings can be adjusted. Instructions for use Oxylog 3000 plus SW 1.n...
The date and time can be set. position and is only displayed at that position. To define the 11 values displayed: Start configuration on page 1/6 and continue through to 6/6. Instructions for use Oxylog 3000 plus SW 1.n...
The selected test remains active until the rotary – RR to 20/min knob is pressed again. – Pmax to 40 mbar – FiO to 40 % These settings are displayed on the screen. To test the buttons: Instructions for use Oxylog 3000 plus SW 1.n...
A maximum of 570 logbook entries can be displayed on the screen. When the capacity of the logbook is reached, the oldest entries are overwritten. Display (example): Instructions for use Oxylog 3000 plus SW 1.n...
– Filter/gas check (Filter/gas check); – Filter check (Filter check); – Gas check (Gas check); – Gas Calibration (Gas Calibration); – Start gas calibration (Start gas calibration); – Reset to factory defaults (Reset to factory defaults) Instructions for use Oxylog 3000 plus SW 1.n...
Page 104
(Zero calibration in progress). 3 Confirm with the rotary knob. The If zero calibration was successful: Oxylog 3000 plus performs the filter check and After approx. 5 seconds the Oxylog 3000 plus displays the message Filter check in progress confirms with the message Zero calibration (Filter check in progress).
Page 105
Configuration The Oxylog 3000 plus displays the alarm message 4 Open the page CO sensor (CO sensor), Filter check failed (Filter check failed). The select the line Filter/gas check (Filter/gas measured CO value is outside the permissible check) and then the line Gas check (Gas tolerance range.
Page 106
CONTINUE (CONTINUE) and confirm with the rotary knob. If the calibration was unsuccessful, the previously valid calibration remains effective. The Oxylog 3000 plus starts the calibration of the CO sensor and displays the progress and For connecting the CO sensor and cuvette, see result of the calibration in the message field.
1 Select the Alarm sequence (Alarm sequence) line and confirm. 2 Select the alarm tone type DRAEGER or IEC by turning the rotary knob and confirm. NOTE The user cannot change the preset alarm tone type. Instructions for use Oxylog 3000 plus SW 1.n...
Configuration Customer service manual For further information on the Oxylog 3000 plus, refer to the Service Manual (can be ordered through DrägerService). Instructions for use Oxylog 3000 plus SW 1.n...
Page 109
......117 Messages in the information field..118 Error messages during the device check . 120 Instructions for use Oxylog 3000 plus SW 1.n...
There are only power supply, or a fully charged approximately 10 minutes of battery must be installed (venti- operating time remaining in the lation stops while installing the internal battery. battery). Instructions for use Oxylog 3000 plus SW 1.n...
Page 111
CO sensor. Perform the CO gas check. sensor? The connector of the CO sensor Reinsert the connector. was removed during operation. The CO sensor has a hardware Replace the CO sensor. failure. Instructions for use Oxylog 3000 plus SW 1.n...
Page 112
Ventilation is continued with last faulty (RR) settings. Check patient condition and ventilation. Contact DrägerService. Control knob Technical defect. Ventilation is continued with last faulty (VT) settings. Check patient condition and ventilation. Contact DrägerService. Instructions for use Oxylog 3000 plus SW 1.n...
Page 113
Call DrägerService. The internal battery is not being Continuous ventilation with this charged due to a device malfunc- device is only possible with an tion. external power source. Call DrägerService. Instructions for use Oxylog 3000 plus SW 1.n...
Page 114
Actual settings will be affected. Call DrägerService. lost in case of a power loss. Loudspeaker Technical defect. To continue ventilation with this inop device, continuously monitor the device functions. Call DrägerService. Instructions for use Oxylog 3000 plus SW 1.n...
Page 115
Check patient condition, check way pressure has been reached. ventilation pattern, correct alarm Patient "fights" the ventilator, limits if necessary. coughing. Breathing hose kinked, or Check breathing circuit, breath- obstructed. ing valve, and tube. Instructions for use Oxylog 3000 plus SW 1.n...
Page 116
Advisory message, is displayed hose 250 mL while using a pediatric continuously when confirmed. breathing hose. Set a lower VT. An incorrect hose connected. Advisory message, is displayed continuously when confirmed. Use another hose. Instructions for use Oxylog 3000 plus SW 1.n...
Additional messages in the alarm message field Message Cause Remedy Self test OK The device has been switched on The message disappears uto- and the selftest completed success- matically after approximately fully. 15 seconds. Instructions for use Oxylog 3000 plus SW 1.n...
(optional)" on knob, the filter check will page 91. start. Cancel by pressing Alarm Reset. Pinsp>=PEEP + 3 mbar PEEP + 3 mbar > Pinsp Set Pinsp > PEEP + 3 mbar. Instructions for use Oxylog 3000 plus SW 1.n...
Page 119
Change in I:E or Ti, in ventila- Te = 9.5 s tion mode PC-BIPAP. I:E= 1 : 1.5 Te = 9.5 s 1) Unit I:E or Ti is configurable. Refer to the "Customer Service Mode" on page 97. Instructions for use Oxylog 3000 plus SW 1.n...
Check measuring lines The flow measuring lines are Connect the flow measuring lines incorrectly connected. correctly. Instructions for use Oxylog 3000 plus SW 1.n...
Page 121
Reprocessing list ..... 129 Assembling parts ..... 129 Instructions for use Oxylog 3000 plus SW 1.n...
(D). 4 Remove the angled connector (A) from the cuvette (C). 1 Disconnect the breathing hose (A) from the expiratory port. 2 Detach the flow measuring lines (B) from the connectors. Instructions for use Oxylog 3000 plus SW 1.n...
Page 123
Risk of CO rebreathing breathing valve. The rubber disk (A) in the housing must not be removed, damaged or bent, otherwise the valve will not work properly and will endanger the patient. Instructions for use Oxylog 3000 plus SW 1.n...
The disposable breathing circuit must not be withstand high temperatures and may be cleaned, disinfected or sterilized: it cannot damaged. withstand high temperatures and may be damaged. Instructions for use Oxylog 3000 plus SW 1.n...
Referring to its reprocessing, this medical device Manual disinfection and simultaneous cleaning belongs to the group of non-critical medical Use disinfectants that are approved nationally devices. and are suitable for the reprocessing procedure being applied. Instructions for use Oxylog 3000 plus SW 1.n...
Neodisher Forte MediClean from Dr. Weigert (for thermal disinfection at 93 °C (199 °F) for – Yeasticidal 10 minutes) – Virucidal or virucidal against enveloped viruses Observe the specifications of the surface disinfectant manufacturers. Instructions for use Oxylog 3000 plus SW 1.n...
Page 127
Observe the applicable country-specific listings for disinfectants. The list of the German Association for Applied Hygiene (Verbund für Angewandte Hygiene VAH) applies in German-speaking countries. Instructions for use Oxylog 3000 plus SW 1.n...
(e.g., disinfectant residues can attack the material more intensely during autoclaving). If signs of wear become visible, such as cracks, deformation, discoloration, peeling, etc., affected accessories must be replaced. Instructions for use Oxylog 3000 plus SW 1.n...
– Connect to the power supply and gas supply, refer to "Assembly" on page 35 for information. – Check readiness for operation, refer to "Preparation" on page 51 for information. Instructions for use Oxylog 3000 plus SW 1.n...
Page 130
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 131
Repair ......134 In case of ventilator failure ... . . 135 Instructions for use Oxylog 3000 plus SW 1.n...
Repeated indicated measures intended to retain the functional integrity of a medical device. Repair Measures intended to restore the functional integrity of a medical device after the failure of a device function Instructions for use Oxylog 3000 plus SW 1.n...
7 Perform a device check according to the instructions for use. 3 Check that the device combination is in good condition: – Labels complete and legible. – No damage. Instructions for use Oxylog 3000 plus SW 1.n...
1) The dust filter can be treated as household waste. 2) Refer to "Technical Data" section for battery operating time. Repair For repairs, Dräger recommends engaging DrägerService and using original Dräger parts. Instructions for use Oxylog 3000 plus SW 1.n...
Never operate a ventilator if it has suffered physical damage or does not seem to operate properly. In this case, always engage factory trained or authorized expert personnel to service the device. Instructions for use Oxylog 3000 plus SW 1.n...
Page 136
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 137
Disposingof the breathing circuits and the cuvettes ......138 Instructions for use Oxylog 3000 plus SW 1.n...
Page 138
Risk of explosion and of chemical burns. Improper handling of batteries can result in explosions and chemical burns. Do not throw batteries into fire. Do not force batteries open. Do not recharge batteries. Instructions for use Oxylog 3000 plus SW 1.n...
Page 139
Electromagnetic environment ... . . 152 Recommended separation distances from wireless communication devices ..152 Instructions for use Oxylog 3000 plus SW 1.n...
5 to 95 % (no condensation) Replaceable battery Temperature –20 to +50 °C (preferred long-time storage tempera- ture <30 °C) Ambient pressure 570 to 1200 hPa Relative humidity 5 to 95 % (no condensation) Instructions for use Oxylog 3000 plus SW 1.n...
±10 Vol%. The actual value depends on the inspiratory flow and mean air- way pressure. Positive end-expiratory pressure PEEP 0 to 20 mbar, no negative pressure. Trigger sensitivity (flow trigger) 1 to 15 L/min Instructions for use Oxylog 3000 plus SW 1.n...
<4 mbar at 30 L/min <2 mbar at 5 L/min Inspiratory and expiratory resistance of the <5 mbar at 30 L/min device with pediatric breathing circuit <2 mbar at 5 L/min Instructions for use Oxylog 3000 plus SW 1.n...
Page 143
The maximum inspiratory flow is reduced to 80 L/min at supply pressures <350 kPa and to 39 L/min at supply pressures <280 kPa. 2) When using an accessory with a female connector, add 2 mL to the breathing circuit dead space. Instructions for use Oxylog 3000 plus SW 1.n...
Accuracy ±1/min Curve display Airway pressure Paw (t) –10 to 100 mbar Flow (t) –120 to 120 L/min –5 to 120 mmHg / –1 to 16 Vol% / –1 to 16 kPa Instructions for use Oxylog 3000 plus SW 1.n...
> 25 seconds above the set value Range of settings 10 to 100/min Leakage Alarm Only with NIV off in VC modes: When VTe <60 % of VTi over the last 25 seconds or 8 respiratory cycles Instructions for use Oxylog 3000 plus SW 1.n...
The specified charging time applies when recharg- ing the battery to a level of approx. 90 % after it has been depleted. Charging of a completely depleted battery is only possible when the ventilator is switched OFF. Instructions for use Oxylog 3000 plus SW 1.n...
Page 147
12 / 24 / 28 VDC; 5 A / 2.5 A / 2.1 A Output 19 V ±0.5 V / 2.6 A The intended use of the DC/DC converter is in vehi- cles. Instructions for use Oxylog 3000 plus SW 1.n...
75 mm (in addition to the width of the basic unit) AC/DC power pack 150 × 37 × 64 mm DC/DC converter 160 × 35 × 60 mm Weight Basic unit without internal Approx. 5.3 kg battery Instructions for use Oxylog 3000 plus SW 1.n...
Page 149
(body floating, defibrillation-proof) reusable), including CO sensor, endotracheal tubes, or masks Protection class against ingress of liquids IPX4, device is splashproof from all directions Protection class of the CO sensor IP64 Defibrillation recovery time Instructions for use Oxylog 3000 plus SW 1.n...
This device was tested for electromagnetic compatibility using accessories from the list of accessories. Other accessories may only be used The EMC conformity of the Oxylog 3000 plus if they do not compromise the electromagnetic includes the use of following external cables, compatibility.
150 kHz to 80 MHz: 3 V, ISM bands: 6 V (IEC 61000-4-6) Electromagnetic fields in the vicinity of wireless Various frequencies from 385 MHz to 5785 MHz: communication devices 9 V/m to 28 V/m Instructions for use Oxylog 3000 plus SW 1.n...
Page 153
Maintain an adequate distance between this device and other medical electrical equip- ment. Instructions for use Oxylog 3000 plus SW 1.n...
Page 154
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 155
Monitoring......164 measurement ..... 164 Instructions for use Oxylog 3000 plus SW 1.n...
This phase, the inspiratory pause, can be identified as the plateau Tplat and is defined as a percentage of the inspiratory time. Instructions for use Oxylog 3000 plus SW 1.n...
Volume Controlled – Synchronized Intermittent respiratory rate, the Oxylog 3000 plus prolongs the Mandatory Ventilation spontaneous breathing time of the subsequent breath by the difference T - thus preventing an...
Page 158
The device then increases the airway pressure up to the preselected pressure Psupp above PEEP, which is adjustable to the condition of the patient. Instructions for use Oxylog 3000 plus SW 1.n...
VT. controlled by the Slope setting. The tidal volume VT results principally from the pressure difference between PEEP and Pinsp and Instructions for use Oxylog 3000 plus SW 1.n...
With AutoFlow, the spontaneous breathing demand. inspiratory flow will adjust a maximum of 3 mbar When the patient inhales, the Oxylog 3000 plus breath to breath. delivers additional inspiratory flow. If the tidal volume VT is reached (inspiratory...
The inspiratory flow rise from the PEEP level to the inspiratory level can be even more closely adapted to the needs of the patient by means of the pressure rise time Slope. Instructions for use Oxylog 3000 plus SW 1.n...
(owing to a long Ti), the cycle time is automatically increased. These rules are also applicable when the ventilator is configured to set I:E instead of Ti. Instructions for use Oxylog 3000 plus SW 1.n...
Oxylog 3000 plus are controlled by the for BTPS, taking into account the ambient microprocessor system via digitized electrical pressure. In this way, the Oxylog 3000 plus meters signals. and measures roughly 10 % less volume in operation with a test lung (dry gas at room temperature).
PEEP, PIP, Pplat, Pmean. The plausibility of this airway pressure measurement on the patient side is monitored by a redundant internal airway pressure measurement in the ventilator via S4 in the inspiratory duct. Instructions for use Oxylog 3000 plus SW 1.n...
Page 165
3 m, 5 pc. TwinStar (filter - HME combination, VentStar Oxylog disposable 57 04 964 50 pieces): breathing circuit for pediatric patients, 1.9 m, 5 pc. 90 mL MP 01 800 Instructions for use Oxylog 3000 plus SW 1.n...
Page 166
57 05 301 Trolley 57 90 150 Options measurement 57 05 331 Real time data export 57 05 332 AutoFlow 57 05 333 inhalation 57 05 329 100 % O 57 05 330 Instructions for use Oxylog 3000 plus SW 1.n...
Page 167
Connecting the gas supply ....41 Contraindications ......14 Instructions for use Oxylog 3000 plus SW 1.n...
Page 168
VC-SIMV......157 Pressure support ....69, 71, 158 Instructions for use Oxylog 3000 plus SW 1.n...
Page 169
VTe ........90 Warning ......2, 86, 88 Instructions for use Oxylog 3000 plus SW 1.n...
Page 170
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 171
Switching on Customer Service Mode ..172 Accessing Customer Service Mode Oxylog 3000 plus SW 1.n ....173 Instructions for use Oxylog 3000 plus SW 1.n...
Page 172
If the cutout containing the access description has been removed, ask your device manager how to make changes in Customer Service Mode. Instructions for use Oxylog 3000 plus SW 1.n...
Page 173
Accessing Customer Service Mode Accessing Customer Service Mode Oxylog 3000 plus SW 1.n Settings in Customer Service Mode 6 Select the required function with the cursor (asterisk). – To select the parameter: turn rotary knob. Oxylog 3000 plus – To activate the parameter: Press the rotary knob.
Page 174
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 175
This page has been left blank intentionally. Instructions for use Oxylog 3000 plus SW 1.n...
Page 176
These instructions for use only apply to Oxylog 3000 plus SW 1.n. with the Serial No.: If no Serial No. has been filled in by Dräger, these instructions for use are provided for general information only and are not intended for use with any specific machine or device.
Need help?
Do you have a question about the Oxylog 3000 plus and is the answer not in the manual?
Questions and answers