Operation
Setting O
and flow in the dialog window
2
The O
and flow can also be set in the Ventilation
2
settings dialog window.
Touch the Ventilation settings... button.
Or
Touch the
(D) button.

Standby mode

Switch to standby mode for the following actions:
– Keep Babylog VN500 ready for operation while
the patient is absent
– Change the therapy type between ventilation
and O
therapy
2
– Change the patient category
– Change the application mode
– Perform the device and breathing circuit check
– Switch off Babylog VN500
WARNING
Ventilation does not take place in standby
mode! The device must only be set to standby
mode when no patient is connected to the de-
vice. Other settings may jeopardize the pa-
tient!

Activating standby mode

1 Touch the Start/ Standby... button in the main
menu bar.
Babylog VN500 opens the Start/Standby dialog
window. The Start/Standby (A) page appears by
default.
102
Switching off O
1 Touch the Start/ Standby... button.
Babylog VN500 opens the Start/Standby dialog
window. The Start/Standby page appears by de-
fault.
2 Touch the Standby button and confirm with the
rotary knob.
Babylog VN500 is in standby mode. The
O
therapy is switched off The therapy type can be
2
switched to ventilation.
Start/Standby
A
2 Touch the Standby (B) button and confirm with
the rotary knob.
The message Standby mode activated is dis-
played in the header bar.
3 Touch the ALARM RESET button in the header
bar and confirm with the rotary knob.
Babylog VN500 is in standby mode. Standby is
displayed in the screen header bar.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
therapy
2
B