Check readiness for operation

The system check consists of the device and
breathing circuit check.

Safety information on the system check

WARNING
The device must not be put into operation if a
fault is discovered during the device and
breathing circuit check! Patient hazard!
CAUTION
The device and breathing circuit check must be
carried out before use on the patient in order to
confirm that Babylog VN500 is operating cor-
rectly.

Starting the system check

The system check is only possible in standby
mode.
1 Touch the Start/ Standby... button in the main
menu bar.
2 Touch the System check (A) tab.
Start/Standby
A
Babylog VN500 displays the date, time and results
of the last system check on the Overview (B) page.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n

Performing a device check

The device check is only possible in standby mode.
Keep test lung ready
Pediatric test lung (8409742) for the pediatric and
neonatal breathing circuit
The test lung must only be inserted into the patient
connector of the Y-piece after instruction by
Babylog VN500.
Starting the device check
Prerequisite: The System check (A) page is
opened.
1 Touch the Device check (B) tab.
Start/Standby
D
B
Babylog VN500 displays the individual test steps in
a list (C). The size of the list depends on the avail-
able applications.
2 Touch the Start (D) button.
3 Confirm with the rotary knob.