Gas Supply Unit Gs500 - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Gas supply unit GS500

Babylog VN500 can be equipped with the GS500
gas supply unit in order to ensure the air supply to
the device in the event of failure of the central gas
supply and during the transportation of patients
within the hospital.
When Babylog VN500 is supplied from the GS500
gas supply unit, the inspiratory O
may deviate from the set FiO
Installing the bacterial filter
CAUTION
In order to protect the patient from contamination
through the aspirated ambient air, an inspiratory
bacterial filter must always be used.
Fit the bacterial filter (A) onto the inspiratory
port.
Using the gas supply unit
Prerequisite: Functionality of the gas supply unit is
activated, see "Device extensions" on page 164.
In the event of failure of the central gas supply, or if
the connectors of the compressed gas hoses be-
come detached from the wall outlets of the central
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
concentration
2
.
2
A
gas supply, Babylog VN500 displays an alarm mes-
sage. The gas supply unit starts the supply of air af-
ter 4 seconds at the latest.
Switching on the gas supply unit for the transporta-
tion of patients within the hospital:
1 Touch the Special procedures... button in the
main menu bar.
2 Touch the Transport (B) tab.
3 Touch the On (C) button.
Special procedures
B
C
D
4 Detach the connectors of the compressed gas
hoses from the wall outlets of the central gas
supply.
If the connectors of the compressed gas hoses
have not been detached within 5 minutes of the gas
supply unit being switched on, Babylog VN500
switches off the gas supply unit.
Switching off the gas supply unit:
5 Touch the Off (D) button.
Additional information
Deactivating functionality of the gas supply unit,
see "Device extensions" on page 164.
Operation
99