Configuration
System settings
The following system settings can be configured:
– Language, date and time
– Units for measured values and settings
– Network and serial interfaces
– Supply units
– Service information
The customized settings are immediately effective.
1 Touch the System setup... button.
2 Touch the System tab.
Babylog VN500 displays the following configurable
settings in an overview:
– Country
– Units
– Interface
– Supply units
– Service
Selecting country-specific settings
Prerequisite: The System (A) page is opened.
1 Touch the Country (B) tab.
System setup
C
E
F
H
I
:
K
Selecting the screen text language
Babylog VN500 is factory set to the customer's own
language. The current language is displayed in the
(C) field.
Selecting a different language:
162
A
D
G
J
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
2 Touch the
Babylog VN500 opens the selection list containing
the available languages.
3 Select the language with the rotary knob and
push to confirm.
Setting the date and time
Babylog VN500 does not change over automatical-
ly between daylight saving time and standard time.
The user must change the time manually. Other-
wise the times will be incorrect in the screen and for
saved values and actions (e.g., in the logbook).
Changing the system time changes the time dis-
played in trends, logbook, alarm history, maneuver
measured values and reference loops. The data
saved up to the change is displayed with the sys-
tem time up till then.
1 Touch the appropriate button:
– Day (E)
– Month (F)
– Year (G)
– Hours (H)
– Minutes (I)
The order of the (E) and (F) buttons varies depend-
ing on language.
B
2 Set the value by turning the rotary knob and
push to confirm.
3 After completing all the settings for the date and
time, touch the Apply (J) button and confirm
with the rotary knob.
(D) button.