Configuring Alarm Settings; Setting Start-Up Values For Alarm Limits - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Configuring alarm settings

The customized settings for the start-up values of
the alarm limits become effective with the admis-
sion of a new patient. The customized alarm tone
settings are effective immediately, depending on
the time of day. The selection of the alarm tone se-
quence is effective immediately.
1 Touch the System setup... button in the main
menu bar.
2 Touch the Alarms tab.
3 Enter password and confirm with Enter.
Babylog VN500 displays the following configurable
alarm settings in the overview:
– Start-up values for alarm limits
– Alarm volume and alarm tone

Setting start-up values for alarm limits

CAUTION
If several devices of the same type are used on a
ward, the alarm defaults must be configured iden-
tically on all devices. Other settings may jeopar-
dize the patient.
Prerequisite: The Alarms (A) page is opened.
1 Touch the Preset limits (B) tab.
System setup
A
C
D
Babylog VN500 displays the start-up values used
for the alarm limits.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
(C)
(D)
2 Touch the relevant button for the alarm limit.
3 Set the value by turning the rotary knob and
push to confirm.
Selecting the factory settings
Touch the Dräger default (E) button and con-
firm with the rotary knob.
B
E
upper alarm limit
lower alarm limit
Configuration
149