Calibrating The O 2 Sensor; Sensor - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Monitoring
Babylog VN500 displays calibration information in
the (C) message field.
At the completion of calibration, the Start button
turns pale green.
Canceling calibration of the neonatal flow sen-
sor
Touch the Cancel (D) button.
Calibrating the O
The O
sensor is calibrated during the device
2
check. Regular calibration during the device check
ensures the specified accuracy. If the O
not calibrated for 3 months, the accuracy of the
O
sensor will be reduced. The parameter field for
2
the O
concentration also displays a question mark
2
in addition to the measured value.
FiO
Vol%
2
22?
After calibration during the device check the sensor
will work again with full accuracy.
CAUTION
If the quality of the oxygen from the central gas
supply is not sufficient, calibrate the O
with calibration gas (100 % O
may result in an incorrect calibration.
Additional information
"Performing a device check" on page 63.
130

sensor

2
sensor is
2
sensor
2
). Otherwise this
2
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
After calibration of the neonatal flow sensor
8 Connect the tube connector.
Setting the flow trigger
Touch the Trigger (E) button.
Babylog VN500 opens the page for setting the flow
trigger. For additional information, see "Additional
settings for ventilation" on page 79.