Dräger Babylog VN500 Disassembly/Assembly And Reprocessing

Dräger Babylog VN500 Disassembly/Assembly And Reprocessing

Neonatal expiratory valve and neonatal flow sensor
Hide thumbs Also See for Babylog VN500:

Advertisement

Quick Links

Dräger Babylog
®
Neonatal Expiratory Valve
and Neonatal Flow Sensor
Disassembly/Assembly and Reprocessing
Disassembly
/Assembly
Open the flap by lifting the lower
edge upwards.
Close the flap.
Disassembly
/Assembly
Turn the locking ring as far as
possible to the left.
Turn the locking ring as
far as it will go to the right
until it clicks into place.
Remove and empty the water
trap container.
Fit the water trap container.
This guideline does not replace or amend the instructions for use for Infinity Acute Care System Workstation Neonatal Care.
The instructions for use and the hospital guidelines and policies always have to be observed.
VN500
Muffler
Pull the muffler as far it will go
to the left.
Push the muffler to the
Neonatal Expiratory Valve
Remove the expiratory valve from
the fitting.
Turn the locking ring as far as
possible to the left and push the
expiratory valve into the fitting.
The muffler is only attached to the Babylog
an 11 mm diameter port. The Babylog
pediatric patient population.
right into the socket.
VN500 is dedicated to neonatal and
®
Remove the muffler from
the socket.
Fit the muffler into the socket.
Remove the diaphragm.
Fit the diaphragm onto the edge
of the expiratory valve housing.
VN500. It is only available with
®

Advertisement

loading

Summary of Contents for Dräger Babylog VN500

  • Page 1 Dräger Babylog VN500 ® Neonatal Expiratory Valve and Neonatal Flow Sensor Disassembly/Assembly and Reprocessing Disassembly /Assembly Muffler Open the flap by lifting the lower Pull the muffler as far it will go Remove the muffler from edge upwards. to the left. the socket.
  • Page 2 Dräger Babylog VN500 ® and Neonatal Expiratory Valve Reprocessing Neonatal Flow Sensor Disassembly/Assembly Reprocessing of Neonatal Expiratory Valve Before manual cleaning prepare cleaning solution. Before and after immersion fit a syringe containing 20 ml of solution to the ejector channel. Inject and extract the solution several times.
  • Page 3 Dräger Babylog® VN500 and Dräger Neonatal Flow Sensor Reprocessing instructions Manual cleaning and disinfection Reprocessing of ISO 15 Neonatal Flow Sensor and Y-piece Neonatal Flow Sensor* At the beginning and end of the Prepare cleaning solution. entire reprocessing procedure vigorously swirl both parts for at least one minute.
  • Page 4 Dräger Babylog VN500 ® and Neonatal Flow Sensor Reprocessing instructions Manual cleaning and disinfection Reprocessing instructions overview: Manual cleaning Machine cleaning Sterilization and disinfection and disinfection housing and sensor separately without sensor sensor and housing connected housing and sensor separately without sensor sensor and housing connected Machine cleaning and disinfection:...