Dräger Babylog
®
Neonatal Expiratory Valve
and Neonatal Flow Sensor
Disassembly/Assembly and Reprocessing
Disassembly
/Assembly
Open the flap by lifting the lower
edge upwards.
Close the flap.
Disassembly
/Assembly
Turn the locking ring as far as
possible to the left.
Turn the locking ring as
far as it will go to the right
until it clicks into place.
Remove and empty the water
trap container.
Fit the water trap container.
This guideline does not replace or amend the instructions for use for Infinity Acute Care System Workstation Neonatal Care.
The instructions for use and the hospital guidelines and policies always have to be observed.
VN500
Muffler
Pull the muffler as far it will go
to the left.
Push the muffler to the
Neonatal Expiratory Valve
Remove the expiratory valve from
the fitting.
Turn the locking ring as far as
possible to the left and push the
expiratory valve into the fitting.
The muffler is only attached to the Babylog
an 11 mm diameter port. The Babylog
pediatric patient population.
right into the socket.
VN500 is dedicated to neonatal and
®
Remove the muffler from
the socket.
Fit the muffler into the socket.
Remove the diaphragm.
Fit the diaphragm onto the edge
of the expiratory valve housing.
VN500. It is only available with
®