Dräger Babylog VN500 Instructions For Use Manual page 228

Infinity acute care system
Table of Contents

Advertisement

Technical Data
Operating data (cont'd)
Oil concentration
Particle size
Gas consumption
Consumption for ventilation
Consumption for pneumatic medication
nebulizer
Automatic gas switch-over
Sound pressure level of device during typical ven-
tilation (measured at operator's position in a free-
field in accordance with ISO 3744 at a distance of
1 m (39 inch) and at a height of 1.5 m (59 inch))
Mean sound pressure level L
Mean sound pressure level L
Dimensions (W x H x D)
Babylog VN500 and Infinity C500
Babylog VN500 and Infinity C500 on trolley
Weight
Babylog VN500 and Infinity C500
Babylog VN500 and Infinity C500
on trolley
PS500
GS500
Maximum load
Trolley
Universal holder with standard rail
Humidifier holder 8416325
Humidifier holder G93111
Electromagnetic compatibility (EMC) (conforming
to European Directive 89/336/EEC)
Classification as per EC Directive 93/42/EEC An-
nex IX
UMDNS code Universal Medical Device Nomenc-
lature System - Nomenclature for medical devices
228
eq(A)
eq(A)
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
3
<0.1 mg/m
Dust-free air (filtered with pore size <1 µm)
Depends on ventilation settings
Medical air or O
max. 2.1 bar (or 210 kPa or
2
30.5 psi), max. 11 L/min
if one gas fails, the device switches to the other
gas.
≤45 dB(A)
≤50 dB(A) with GS500
≤45 dB (A) for conventional ventilation
≤50 dB (A) for HFO
420 mm x 685 mm x 410 mm
(16.5 inch x 27.0 inch x 16.1 inch)
577 mm x 1400 mm x 677 mm
(22.7 inch x 55.1 inch x 26.7 inch)
approx. 25 kg (55.1 lbs)
approx. 59 kg (130 lbs)
approx. 25 kg (55.1 lbs)
approx. 10 kg (22 lbs)
100 kg (220.5 lbs)
10 kg (22.1 lbs)
10 kg (22.1 lbs)
8 kg (17.6 lbs)
tested in according to EN 60601-1-2
II b
17-429